Abstract

To achieve accurate results in nanocalorimetry a detailed analysis and understanding of the behavior of the calorimetric system is required. There are especially two system-related aspects that should be taken in consideration: the properties of the empty cell and the effect of the thermal link between sample and cell. Here we study these two aspects for a membrane-based system where heater and thermometer are both in good contact with each other and the center of the membrane. Practical, analytical expressions for describing the frequency dependence of heat capacity, thermal conductance, and temperature oscillation of the system are formulated and compared with measurements and numerical simulations. We finally discuss the experimental conditions for an optimal working frequency, where high resolution and good absolute accuracy are combined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.