Abstract

IntroductionDuring adrenal venous sampling (AVS) procedure, radiologists administer a contrast agent via the catheter to visualize the proper catheter position.Materials and methodsA patient with primary aldosteronism diagnostic-hypothesis was admitted for AVS. A venogram was performed to confirm the catheter’s position with 2mL of Iopamidol 300 mg/mL. Samples were collected with syringe connected to a hydrophilic coated catheter by low-pressure aspiration from each of the four collection sites: inferior vena cava in the suprarenal portion, inferior vena cava in the infrarenal portion, left adrenal vein, and right adrenal vein; then immediately transferred from syringe to tubes with gel separator. All tubes were centrifuged at 1200 x g for 10 minutes.ResultsAt the end of centrifugation process, primary blood tubes containing blood from inferior vena cava and left adrenal vein exhibited the standard gel separator barrier, while tubes from right adrenal vein showed abnormal flotation of gel separator. The radiologist confirmed the usage of 2.6 mL instead of 2.0 mL of Iopamidol 300 mg/mL. This iodinated contrast media, with 1.33 g/cm3 of density, was used close to the right adrenal vein due to some difficulty to access it.ConclusionThe abnormal flotation of gel separator in samples taken from right adrenal vein can be explained by the usage of the iodinated contrast media. We suggest using plain-tubes (without gel separator) for AVS in order to avoid preanalytical nonconformities. Moreover, a blood volume equivalent to twice the catheter extension should be discarded to eliminate residual contrast media before collection of samples for laboratory assays.

Highlights

  • During adrenal venous sampling (AVS) procedure, radiologists administer a contrast agent via the catheter to visualize the proper catheter position

  • The radiologist confirmed the usage of 2.6 mL instead of 2.0 mL of Iopamidol 300 mg/mL

  • The abnormal flotation of gel separator in samples taken from right adrenal vein can be explained by the usage of the iodinated contrast media

Read more

Summary

Introduction

During adrenal venous sampling (AVS) procedure, radiologists administer a contrast agent via the catheter to visualize the proper catheter position. The simultaneous laboratory measurement of cortisol concentrations during catheterization in samples from the adrenal vein allows a correction for eventual dilution, and is an objective evidence of the proper cannulation (e.g., by selectivity index). Laboratory should report selectivity index as soon as possible (e.g., within less than half hour). This procedure allows any improperly collected adrenal samples to be immediately re-collected, reducing the frequency of repeat procedures and thereby reducing the cost as well as the discomfort to the patient and avoiding delay in diagnosis [3]

Methods
Results
Discussion
Conclusion