Abstract
The preparation and performance of a weak cation-exchange stationary phase for Open Tubular Liquid Chromatography (OT-LC) was investigated. The stationary phase was prepared in 5.4 μm I.D. fused silica capillaries byin situ photopolymerization of a mixture of silicon acrylate and acrylic acid. The influence of pH, counter ion concentration and organic modifier concentration of the mobile phase on the retention was studied with catecholamines as test solutes using LIF detection. Other biological amines like amino acids, small peptides and nucleic acid derivatives could be separated on this stationary phase as well. The kinetic performance of the stationary phase was studied with several cations and neutral solutes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.