A gas analysis apparatus is described consisting of a metal vacuum system which includes a mass spectrometer, connected by a capillary leak to a glass vacuum system which can be readily modified to suit any particular problem. This arrangement provides for a wide variety of experiments, examples of which are the analysis of small quantities of gas such as occur in bubbles in glass, desorption studies and vacuum fusion analysis. The minimum quantity of gas that is detectable with this apparatus is approximately 5 1. μtorr. The mass spectrometer may also be used as a leak detector with a detection limit of 0.1 1. ntorr sec-1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.