Abstract

Our aim was to investigate the repeatability and the accuracy of the measured MP at different PEEP values and tidal volume compared with the calculated MP. MP was measured and calculated in sedated and paralyzed ARDS patients at low and high tidal volume, at 5-10-15 cmH2O of PEEP both in volume and pressure-controlled ventilation. The same measurements were performed twice. Fifty ARDS patients were enrolled. MP was measured and calculated for a total of 300 measurements. The bias and limits of agreement were 0.38 from -1.31 to 2.0J/min. The measured and calculated MP were similar in each ventilatory condition. The mechanical power measured by a new automatic real time system implemented in a mechanical ventilator was repeatable and accurate compared with the computed one.