Abstract
An eco-friendly liquid chromatography method was developed for determining imatinib mesylate in pharmaceuticals. During pre-development studies, Using a greener mobile phase and a combination of qualitative and semi-quantitative green metrics techniques, the environmental sustainability of the procedure was maintained. Methanol:water (pH maintained at 3.5 using orthophosphoric acid) at 75:25%, v/v flowing at 1 mL/min was the final mobile phase used on a C-18 column. The diode array detector detected imatinib at 262 nm. Method specificity, accuracy, precision, selectivity and system suitability were tested over a linear concentration range of 5-120 μg/mL of imatinib. The limit of detection and limit of quantitation values were 1.5 and 5.0 μg/mL, respectively. In addition, the factorial design showed the procedure to be reliable with minimum testing. Finally, the method recovered the maximum (> 99%) of the analyte from the dosage form. Overall, the present method is suitable for routine applications in quality control of dosage forms of imatinib.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.