Abstract

A two-way Josephson voltage standard (JVS) direct comparison between the National Institute of Standards and Technology (NIST) and the National Research Council (NRC) has been conducted. The process consists of two comparisons: first, using the NRC JVS with the NRC's measuring system (hardware and software) to measure the 10 V provided by the NIST JVS and then using the NIST JVS measuring system to measure the 10 V provided by the NRC JVS. The results of the two comparisons are in agreement to within 0.7 nV, and their mean indicates that the difference between the two JVSs at 10 V is -0.28 nV, with a pooled combined uncertainty of 2.07 nV (k = 2) or a relative uncertainty of 2.1 parts in 1010.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.