Abstract

A parallel plate avalanche chamber specially suited for the high resolution detection of minimum ionizing particles (m.i.p.) is presented. The anode is made of a thin germanium layer with a sheet resistivity > 1 MΩ/p[ while the cathode is made of a nickel mesh having 600 line pairs/in. A chess board of pads placed behind the anode plane is used to obtain the positional information. A 100% detection efficiency, a 40 ns (fwhm) time resolution and a spatial resolution better than 140 μm (fwhm) for both coordinates have been measured.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.