Abstract
We describe a procedure for enzyme immunoassay of theophylline (1,3-dimethylxanthine) in which all phases of the assay are totally automated in a Kinetic Analyzer (KA-150). This system permits assay of 75 10-mul samples per hour, with results available at 30-s intervals after initial sample preparation and preincubation. We compared results for 138 clinical samples by an ultraviolet method (x) and the present method (y). The slope of the comparison curve was 0.902, the y-intercept 0.402, and the correlation coefficient 0.984. The coefficient of variation for samples run in duplicate on the same day was 4.9%; it was 8.1% for samples run on different days. Specificity, sensitivity, simplicity, speed, and small reagent requirement all make this an attractive alternative to chromatographic procedures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.