Abstract

Initial results are reported of a Polish-Finnish project to verify electron dose distributions calculated by treatment planning systems (TPSs), CadPlan v.6.3.2 and Theraplan v.3.5, which use different electron beam dose distribution algorithms. Treatment of gross tumour volumes representing lung and parotid cancer was simulated in an Alderson anthropomorphic phantom with thermoluminescent detectors (TLDs) (Li(2)B(4)O(7):Mn,Si) placed at selected measurement points inside its volume. The observed discrepancy between relative values of dose calculated and measured by TLDs at each of the measurement points and those calculated by the different TPSs at the same points is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.