Abstract
A test of planar spark counters (PSCs) at the PEP electron-positron storage ring showed the following. PSCs can be used under harsh experimental conditions without long term degradation of resolution. On-line time-of-flight resolution were below 200 ps, coordinate resolutions better than 4 mm, both limited to these values by the resistivity of the semiconducting glass and the electronics used. The best single counter time resolution under realistic conditions at the storage ring was 76 ps. Pulse height was a good indicator of the number of particles striking a counter. Subsequent measurements using cosmic rays gave a single counter time resolution of 50 ps.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods In Physics Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.