Abstract
Motivation of this study is to check the sensitivity of dosimetric tool gamma with 2D detector array combination when unexpected errors occur while transferring intensity-modulated radiation therapy treatment plans from planning system to treatment unit. This study consists of 17 head and neck cancer patient's treatment plans. Nine types of verification plans are created for all 17 clinically approved treatment plans by consecutively deleting different segments (up to eight) one by one from each field of the plan. Decrement factor (χ) is introduced in our study which illustrated the degree of decay of gamma passing rate when intentional errors are introduced. We analyzed the data by two different methods-one without selecting the region of interest (ROI) in dose distributions and the other by selecting the region of interest. By linear regression, the absolute value of slopes is 0.025, 0.024 and 0.015 without ROI and 0.030, 0.027 and 0.015 with ROI for 2%/2mm, 3%/3mm and 5%/5mm criteria, respectively. The higher absolute value of the fitted slope indicates the higher sensitivity of this method to identify erroneous plan in treatment unit. The threshold value for 2%/2mm equivalent to 95% passing criteria in 3%/3mm used in clinical practice is obtained as 83.44%. The 2D detector array with dosimetric tool gamma is less sensitive in detecting errors when unprecedented errors of segment deletion occur within the treatment plans.