The aim of the current study was to test the technical and clinical feasibility of a robotic system and investigate its potential in the surgical repair of perforated Schneiderian membranes using an ex-vivo porcine model. Eight pig heads were operated conventionally via a surgical loop and eight pig heads with the surgical robot “Symani® Surgical System” (Medical Microinstruments, Inc., Pisa, Italy). On each specimen, the Schneiderian membrane was incised over a length of 0.7 mm resembling a perforation. Operation time, the maximum sinusoidal pressure, the course of the pressure and the filling volume were measured. Additionally, adaptation of the wound edges has been detected via scanning electron microscopy. There were no significant differences for the pressure maximum (p = 0.528), for the time until the pressure maximum was reached (p = 0.528), or for the maximum filling volume (p = 0.674). The time needed for the suturing of the membrane via robotic surgery was significantly longer (p < 0.001). However, the scanning electron microscope revealed a better adaptation of the wound edges with robotic surgery. The technical feasibility of robot-assisted suturing of Schneiderian membrane laceration using the robotic system has been confirmed for the first time. No differences considering the pressure resistance compared to the conventional repair could be observed, but advantages in wound adaptation could be found with an electron microscope. Regarding the material and training costs and limited indications spectrum, robotic surgery systems still might not present financially feasible options in the daily dental practice yet.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call