Abstract

Errors of the order of a few per cent may occur in trace element measurements because of changes in the Ge(Li) detector resolution caused by count rate; errors appear when photopeak intensities are measured by summing a fixed number of data channels. A simple correction method based on a gaussian approximation is proposed. The problem is discussed in the context of trace element measurements in archaeological material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.