Abstract

Large volume Ge(Li) detectors have been used in a γ-γ coincidence arrangement employing two Ge(Li) detectors. The first part of this paper reports the results of studies of the timing properties of planar and coaxial diodes and the time resolution curves that can be obtained. For two cylindrical coaxial diodes the full-width at half-maximum of the time resolution curve was 14 nsec for 511 keV γ-rays. The second part of the paper describes the application of the detectors in γ-γ coincidence experiments. The coincidence counting rates are limited by the total singles counting rates that can be tolerated without excessive loss of energy resolution. Results are shown illustrating the application of a Ge(Li)-Ge(Li) coincidence arrangement to a complex γ-ray spectrum and the ability of this technique to reveal new information on γ-γ cascades.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.