Abstract

Capillary gas chromatography with nitrogen-specific detection allows rapid screening of numerous drugs of toxicological interest. However, for accurate identification of individual peaks, the system must be well calibrated, e.g., through the use of retention indices (RI). To overcome problems associated with the use of RI's based on homologous series determined with nitrogen-specific detectors, we have developed an RI reference system based on molecular masses and retention times of nitrogen-containing compounds. The standards chosen are easily available in highly purified form and can be detected by the unmodified nitrogen-specific detector. By using temperature programming, we can obtain a linear relationship between the molecular masses of standards and their retention times. Used in conjunction with microcomputer data handling, this screening system is rugged and reliable, operating 22 h per day. In the past two years, we have screened greater than 3000 samples (blood, serum, urine, gastric lavage) without major problems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.