Abstract

The present paper describes the design and performance of an attachment for the Cary model 14 recording spectrophotometer, which permits the accurate recording of absorption spectra in small areas. A special compartment built into the light path of the spectrophotometer holds a low-magnification microscope (“macroscope”) consisting of two opposed quartz condensers, with which measurements can be made between 300 and 700 mμ in fields 0.1 to 1 mm in diameter. This can be replaced by a conventional microscope with which spectra can be measured from about 350–650 mμ in fields as small as 4 μ in diameter. These arrangements have been used to measure the absorption spectra of visual pigments in situ. With the macroscope, such measurements have been made in small areas of surviving retinas; and with the microscope, they have been made in single isolated outer segments of rods. The present paper contains examples of each type of measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.