Abstract
Two kinds of quantum Hall devices with different working magnetic fields are fabricated and compared with the LEP (Laboratoires d’Electronique Philips) device, which is currently used in the primary resistance standard system of the National Institute of Metrology (NIM) of China. The comparison is made by calibrating the same 1 Ω standard resistor made by the National Measurement Laboratory of Australia using a Tinsley 100 Ω resistor as the intermediate standard and a cryogenic current comparator. The calibrated values from the NIM device with high working magnetic field and those from the LEP device agreed to within −0.69 × 10−9 with an uncertainty of 4.9 × 10−9. The values from the NIM device with low working magnetic field and those from BIPM-2 agreed to within 2.5 × 10−9 with an uncertainty of 7.1 × 10−9.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.