Abstract

Pfund’s combination of the McLeod and Pirani pressure gages was studied with a number of gases. The calibration curves of pressure against galvanometer reading were found to be straight lines and, except for hydrogen, nearly interchangeable for all the gases used. By using a galvanometer of high sensitivity and introducing a compensating filament, the sensitivity of the gage was increased until a change in pressure of 1.5×10−9 mm of mercury would cause a change in deflection of 1 cm on the galvanometer scale.Brief comments are made on the release of carbon monoxide and air from coconut charcoal and on the adsorption of air at low pressure by silica gel.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.