Abstract
A canine total hip arthroplasty model was used to examine micromotion and bone apposition to a proximally two-thirds madreporic-surfaced femoral prosthesis. Micromotion was also measured following initial press-fit implantation into canine cadaveric femora. After initial press-fit fixation and either 6 or 24 months of biologic fixation, micromotion was less than 23 μm in the proximal and midstem regions, a magnitude consistent with bone apposition. Bone apposition was greatest near the junction of the madreporic and smooth surfaces and was not significantly different between 6 and 24 months (51% at 6 months and 47% at 24 months). The quality of the interface tissue appears to be conducive to long-term fixation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.