Abstract

The proximity of the lens to the retina makes the treatment of retinoblastoma a challenge for external beam radiation therapy. The approximately 1 mm separation between the posterior edge of the lens and the anterior region of the retina causes a trade-off between coverage of the entire retina and excessive dose to the lens. A stereotactic, LINAC based, lens sparing technique for treating retinoblastoma is presented. The technique uses noncoplanar arcs with the lens at isocenter. A special noncircular collimator blocks the lens but it also causes the dose distribution to vary across the retina. A fluence modulation filter is used to reduce the dose inhomogeneity across the target. The resulting dose distribution is roughly hemispheric, providing both anterior coverage of the retina and lens blocking unlike conventional techniques. The method used to develop the collimator and filter assembly is presented. Dosimetry of the assembly was carried out using radiochromic film, and the results were entered in a treatment planning system. The dose distribution as measured in a phantom is provided and compared to calculations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.