Abstract
A pulsed light emitting diode (LED) of constant light output is used as a reference light source to stabilize the amplification of a 4″ø × 2″ NE2213 scintillation detector against long term drifts, temperature and count rate variations. The detector pulse height has been kept constant for several days to within ±0.5% [±1.0%] for temperatures between (20±1)°C[(20±5)°C] and count rates up to 20 kHz. The LED is rigidly coupled to the scintillator cell. The whole stabilization system including two coupled regulation loops is housed in a 2 12 NIM case. The electronic circuitry is presented in principle and in detail, and detector test results are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods In Physics Research
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.