Abstract

A single-column solid-phase extraction procedure was developed for the screening of acidic, neutral, and basic drugs from plasma. The recoveries of all 25 tested drugs exceeded 82%. After the plasma had been diluted with phosphate buffer (pH 6.0), the drugs were extracted using a single Bond Elut Certify column. The acidic and most of the neutral drugs were eluted by acetone/chloroform (1:1) and the basic drugs were eluted by 2% ammoniated ethyl acetate. Some neutral drugs appeared in both fractions. The two fractions were collected separately and evaporated until approximately 100 microL of solvent remained in the tube. Both fractions were analyzed separately on a gas chromatograph equipped with a wide-bore capillary column and a flame ionization detector. The procedure could also be used for urine samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call