Abstract

We describe the suitability of the Amicon MPS-1 centrifugal ultrafiltration device and the YMB membrane for measuring free cortisol in serum. The method combines two independent assays: total cortisol and the ultrafiltrate fraction of added [ 3H]cortisol. The unbound fraction is determined in 0.25–0.30 ml of ultrafiltrate collected from 0.6 to 1 ml of serum that has been equilibrated with [ 3H]cortisol at 37°C for 20 min. The assay is rapid (less than 1 h), practical (no more than 0.6 ml of serum is necessary) and repeatable (CV: 3.8% within-assay and 12.2% in different assays). Error introduced in free cortisol measurement due to dilution effects in dialysis is systematically defined, and the effect of tracer purity on the ultrafiltration method is examined. Dialyzed sera from normal men and women, from patients with Cushing's disease and adrenal insufficiency, and from pregnant women gave ultrafiltration results that accurately duplicated those obtained by previous dialysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.