Abstract

Methods are presented for rapid determination of relative efficiencies of Faraday cups in a multicollector array with movable cups and of Daly or electron multiplier detector dead time and gain values. The Faraday calibration approach is based on measuring the same isotopic ratio in two sequences with different collector configurations having one collector in common. Changes in thermal fractionation during the two measurement sequences are monitored using the same two collectors. Since the relative efficiencies are determined by measuring the same ratio in different cups corrected for time-dependent changes in fractionation, it is unnecessary to use a standard of known composition and calculations are simple, not requiring the solution of multiple equations. Determination of dead time and gain values for a Daly detector are shown based on multidynamic measurements of masses 207, 206, and 208 from Pb standard SRM982 in two sequences consisting of L1-Daly-H1 and Daly-H1-H2. This provides two equations that can be solved for Daly dead time and gain. This method uses static measurements and is therefore insensitive to signal instability. It also does not require use of a standard of known isotopic composition. The potential of using known cup efficiencies to help determine absolute isotopic abundances is discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.