Abstract

A simple valid in situ relative intensity calibration technique for soft x-ray film is described. This is based on film exposure measurements of the uniform line-shaped distributed soft x-ray monochromatic irradiation transmitted through a step-wedge absorption filter. Fitting the calibration data with Henke’s semiempirical equation for thick-emulsion film, the characteristic curves for Shanghai 5F soft x-ray film without supercoat (SIOM-5FW) have been obtained in the wavelength region from 50 to 80 Å.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.