Abstract
A direct, inexpensive and solventless method based on headspace extraction coupled to gas chromatography-flame ionization detection (GC-FID) was developed for determination of benzene, toluene, ethylbenzene and xylenes (BTEX). In this method, volatile organic compounds are separated from the matrix and preconcentrated in the headspace of the sample solution. Then, the headspace vapor is transferred into the GC injector using a fused silica capillary tube. The parameters affecting the headspace extraction such as extraction time, extraction temperature, stirring rate, sample volume, transfer line tubing diameter, transfer line depth in the injector and salt addition were carefully investigated and optimized. The limits of detection (LODs) for the studied compounds ranged from 0.1 to 0.3 ng mL−1 for 15 mL sample solution with the relative standard deviations (RSDs) ranging from 3.34 to 7.18% (n = 5). Finally, the optimized method was successfully applied to water and wastewater as real samples.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.