Abstract

A new electroanalytical method for vitamin B1 determination, based on adsorptive chronopotentiometric stripping analysis and non-specific adsorption onto mercury film electrode, was developed and validated. Stripping chronopotentiograms showed a single well-defined oxidation wave corresponding to vitamin B1 at about − 0.43 V in citrate buffer pH 6.0. The most important experimental factors affecting the monitored electroanalytical response of vitamin B1 were investigated and optimised. Under the optimal experimental conditions, linear response of vitamin B1 was obtained in the concentration range of 5–50 mg dm−3, with the achieved limit of detection of 1.64 mg dm−3, and the limit of quantitation of 4.97 mg dm−3. A mean recovery of 97.1% and relative standard deviations of 3.75% were achieved. The developed electrochemical procedure was successfully applied for the determination of vitamin B1 in pharmaceutical products. The results of the proposed method were in good agreement with those obtained by parallel HPLC analyses, confirming an accuracy of the developed method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.