Abstract

A sensitive enzyme immunoassay for the measurement of insulin in human sera on microtiter plates was established. The assay is based on the sandwich technique with guinea pig anti-insulin IgG adsorbed at microtiter plate wells, human insulin as standard and the same anti-insulin IgG labeled with horseradish peroxidase. Standards used cover a range from 0 to 1200 pmol/l with a detection limit of 10 pmol/l. Coefficients of variation between 3-7% for intraassay precision and 5-11% for interassay precision were obtained over the concentration range of 80-1000 pmol/l. The correlation of EIA-data with those of a commercially available double antibody radioimmunoassay (r = 0.98) could be expressed by the equation: EIA = 0.97 RIA - 57 pmol/l. Normal fasting serum insulin concentrations in healthy subjects ranged from 11-165 pmol/l. In subjects with potentially diminished basal values concentrations of 10-79 pmol/l were determined. The insulin response in oral glucose tolerance tests of children was discussed, who had a constitutional tall stature or Turner's syndrome, respectively.