Abstract

We describe an enzyme-linked two-site immunoassay for quantitation of intact insulin in human serum and plasma. The method uses two murine monoclonal antibodies that bind to two different epitopes on the insulin molecule. The immunoassay is specific. Human proinsulin is not bound by the antibodies, and the binding of partially processed proinsulin intermediates is believed to be of minor clinical importance. The relative response of human, bovine, and porcine insulin is 1, 1, and 3, respectively. The assay is sensitive (detection limit 5 pmol/L), accurate (101% recovery with 50 pmol/L insulin added to samples, 95% with 100 pmol/L, and 89% with 300 pmol/L), and fast (results within 3 h), and has a high analytical capacity (done in microtiter plates). The working assay range selected is 5-600 pmol/L, corresponding to a clinically useful range. Because of its specificity, this two-site immunoassay gives results that are lower than those obtained by using a competitive radioimmunoassay, both in normal individuals and in non-insulin-dependent diabetics.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.