Abstract

A semi-empirical calibration method for single-layer thin films is described which is based on measured standards and cross-sections generated by the National Institute of Standards and Technology database XGAM. A Fortran program, XRFCAL, fits the ratio of measured sensitivity to the relative excitation and detection efficiency to a second-order polynomial in atomic number. The approach is intended not only to provide sensitivities for elements that have no reliable standards but also to improve on the calibration for elements that do. It is applicable only to K-line radiation and silicon detector energy-dispersive XRF spectrometers with secondary or monochromatic excitation. Copyright © 1999 John Wiley & Sons, Ltd.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.