Abstract

To improve the ability of an external cavity laser (ECL) biosensor to more easily distinguish true signals caused by biomolecular binding from a variety of sources of background noise, two photonic crystal (PC) resonant reflectors were incorporated into a single flow cell, with one of the PCs performing the detection function and the other one serving as a reference sensor. The ECL-based sensor system simultaneously emits at two distinct wavelengths corresponding to two different longitudinal cavity modes selected by the sensing and reference PC reflectors. The surface of the sensing PC filter was functionalized by a biomolecule recognition layer and exhibited narrowband reflection with the peak reflection wavelength at 856 nm. The reference PC was untreated and had the peak reflection wavelength at 859 nm. The PCs were bond to the upper and lower surfaces of a thin chamber frame, forming a flow cell. Utilizing the reference external cavity mode, the dual-mode ECL sensor system eliminated common-mode noise sources, including thermal drift, refractive index variations of the analyte solution, and nonspecific biomolecule binding.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.