Abstract

In proton radiotherapy, a round dose profile at the isocenter is beneficial to the treatment planning in the active scanning mode. Based on the theory of the round-beam method, a collimator was designed to be placed at the entrance of the gantry. A simulation of Monte Carlo method was done to validate the design scheme. With the constraining function of the collimator, the dose profile at the isocenter would maintain a round shape despite of an initial asymmetric phase space in extraction of synchrotron. Furthermore, the shape of dose profile is irrelevant to the rotating angle of the gantry. The mechanical error caused by the axis deviation of the gantry could also be partly eliminated with the collimator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.