Abstract

A hardware linear interpolator with logarithmic and exponential curve slope correction is described in this paper allowing real time linearity improvement of low resolution sensors like the miniaturized, wearable or implantable ones used for health monitoring. A compression is performed by the interpolator when the input signal is sparse or changes with low frequency. A signal distorted by the digitization process can be approximated even when its available samples are in nonuniform distances. Multiple interpolators with identical or increasing resolution can be connected in series. A three-stage interpolator with 9-bit input and 12-bit output typical resolution is implemented using 14% of the Logic Elements of an Altera Cyclone III EP3C25N Field Programmable Gate Array and is tested using an ADC developed by the author. An improvement close to 100% has been measured at the Signal to Noise and Distortion Ratio (SNDR) and at the Spurious Free Dynamic Range (SFDR) of the ADC when a 2MHz sinusoidal input was used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.