Abstract
A reversed-phase liquid chromatographic method has been developed and validated for the determination of the polydimethylsiloxane (PDMS) component of Simethicone, which is used as an anti-foaming agent in pharmaceutical formulations. The method involves acidification to neutralise antacid components of the formulation, then a single extraction of the PDMS with dichloromethane. This is followed by separation with a reversed-phase column using an acetonitrile–chloroform solvent gradient, and quantification by an evaporative light scattering detector. An assay precision of 3% was achieved in intraday and interday determinations. No interference was found from the aluminium and magnesium hydroxide components of antacid formulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.