Abstract

The focusing properties of an inclined-plane electron spectrometer have been measured with a narrow-slit and intrinsic-germanium detector. The aberrational width of 3.2 ± 1.0 keV obtained from this measurement is in agreement with ray-trace calculations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.