Abstract

ABSTRACT Purpose Braided flow diverters (FD) are highly sophisticated, delicate, and intricate mechanical devices used to treat intracranial aneurysms. Testing such devices in vitro, however, remains an unsolved challenge. Here, we evaluate methods to measure flow, design and mechanical properties in vitro. Methods Flow properties, cell porosity, pore density, and cell area were evaluated under geometrically realistic conditions by placing FDs in patient-derived, 3D-printed models of human vasculature. 4D flow MRI was used to measure fluid dynamics. Laser microscopy was used to measure the design properties of the FDs. New methods were developed to investigate the bending, circumferential, and longitudinal radial force of the FDs continuously over varying diameters. Results The placement and flow properties of the FD in the vasculature models were successfully measured by MRI, although artifacts occurred. Likewise, the porosity, pore density, and cell area were successfully measured inside of the models using a laser microscope. The newly developed mechanical methods allowed to measure the indicated forces – to our knowledge for the first time - continuously. Conclusion Modern and specifically tailored techniques, some of which were presented here for the first time, allow detailed insights into the flow, design, and mechanical properties of braided flow diverter stents

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.