Abstract

This work describes a rapid method to correct the two-fold astigmatism of transmission electron microscope (TEM) objective lens employing caustic curve when no objective aperture is inserted. The method makes use of rounding the caustic curve via the objective lens stigmators after the condenser lens astigmatism has been corrected. It has many advantages over other methods, it is fast, straightforward, and does not need holes or an amorphous material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.