Abstract

A rapid, sensitive and reproducible HPLC method was developed and validated for the analysis of pantoprazole (CAS 102625-70-7) in human plasma. The separation was achieved on a monolithic silica column using acetonitrile-potassium dihydrogen phosphate buffer (25:75,v/v), pH 3.0, as the mobile phase at a flow rate of 1.5 ml min(-1). The wavelength was set at 290 nm. The assay enables the measurement of pantoprazole for therapeutic drug monitoring with a minimum quantification limit of 20 ng ml(-1). The method involves a simple protein precipitation procedure. Analyticil recovery was complete. The calibration curve was linear over the concentration range 20-3500 ng ml(-1) The coefficients of variation forthe inter-day and intra-day assay were found to be less than 7%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.