Abstract
The description of the injector system, the calibration results and calibration factors are discussed. The ratio static efficiency (standard beta disc sources at fixed geometry) to dynamic efficiency was dependent on the energy and the geometry. Four our geometry these ratios are constant (4580 and 5630 respectively for 133 Xe and 11 C). By making some assumptions, calibration factors for other energies can be determined by a measurement of the static efficiency. Lastly, the total uncertainties in the calibration factor for a determined activity were estimated to be less than 10% and approximately 20% for determining air activity concentration
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.