Abstract
ObjectiveIt is unclear whether supplemental oxygen and noninvasive ventilation respiratory support devices increase the dispersion of potentially infectious bioaerosols in a pressurized air medical cabin. This study quantitatively compared particle dispersion from respiratory support modalities in an air medical cabin during flight. MethodsDispersion was measured in a fixed wing air ambulance during flight with a breathing medical mannequin simulator exhaling nebulized saline from the lower respiratory tract with the following respiratory support modalities: a nasal cannula with a surgical mask, high-flow nasal oxygen (HFNO) with a surgical mask, and noninvasive bilevel positive airway pressure (BiPAP) ventilation. ResultsNasal cannula oxygen with a surgical mask was associated with the highest particle concentrations. In the absence of mask seal leaks, BiPAP was associated with 1 order of magnitude lower particle concentration compared with a nasal cannula with a surgical mask. Particle concentrations associated with HFNO with a surgical mask were lower than a nasal cannula with a surgical mask but higher than BiPAP. ConclusionsParticle dispersion associated with the use of BiPAP and HFNO with a surgical mask is lower than nasal cannula oxygen with a surgical mask. These findings may assist air medical organizations with operational decisions where little data exist about respiratory particle dispersion.