Abstract

Quantitative measurement of marker expression in diseased cells is still a topic of considerable interest and different methodologies are currently under intense scrutiny. This work aims at developing an in vitro diagnostic method based on the release of paramagnetic species from relaxometrically “silent” liposomes operated by the action of a phospholipase A2 (PLA2) previously targeted to the epitope of interest. The released paramagnetic species causes an increase of the longitudinal water proton relaxation rate proportional to the number of PLA2 bound to the cell outer surface. The sensitivity of the herein proposed method, named R-ELISA, was attempted in the detection of folate receptor expression on human ovarian cancer cells by functionalizing PLA2 with folic acid. Receptor/cell number of 8.3×105 has been measured on IGROV-1 cells. The R-ELISA assay can detect nanomolar cell suspension receptor concentrations and has been validated by well-established spectrofluorimetric procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.