Abstract
A prosthesis constructed with a fatigued steel helix encased in a silicone rubber shield was used to band the main pulmonary artery in 10 dogs. After a mean duration of 138 days the banded site was dilated with a 20 mm diameter angioplasty catheter. This dilatation produced a mean increase of 44.3% in the cross-sectional area. A further mean increase of 2.2% in the cross-sectional area was measured 137 days after the dilatation. In five uncomplicated experiments a second dilatation was performed with a 23 mm diameter angioplasty catheter after a mean interval of 140 days. The second dilatation produced a further 21% increase in the cross-sectional area. In the five experiments in which two dilatations were performed, there was a total increase in the mean cross-sectional area of 94% produced 273 days after banding. This prosthesis maintains banding of the main pulmonary artery and can be serially dilated by balloon angioplasty.