Abstract

In this study, neat PCL, neat PLA and PLA/PCL (50/50) based tubular surfaces are produced by electrospinning to simulate the native blood vessel structure and to investigate the effects of both graft material and fiber orientation on burst strength. The burst pressure values of these vascular graft structures that designed with both randomly oriented fibers and oriented fibers, measured by a custom- burst pressure tester, and the results are compared. The results show that fiber orientation have a great influence on burst pressure, regardless of the type of biomaterial. It is determined that grafts with oriented fibers have at least twice the burst strength than those with random fibers. The findings indicate that changing the graft material has also an effect on burst strength. When the results are analyzed by polymer type, although the PLA100_O sample has the highest burst strength among all oriented fiber sample groups, it is better to determine the vascular graft candidate by taking into account radial elasticity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.