Abstract
Background The transobturator posterior anal sling (TOPAS) system is a posterior anal sling that is a minimally invasive, self-fixating polypropylene mesh intended to treat fecal incontinence (FI) in women who have failed conservative therapy. Objective We are reporting 1-year outcome in a prospective, multicenter study under investigational device exemption, evaluating this new treatment modality. Study Design A total of 152 women were implanted with the TOPAS system at 14 centers in the United States. FI was assessed preoperatively and at the 12-month follow up with a 14-day bowel diary, Cleveland Clinic Incontinence Scores, and FI Quality of Life questionnaires. Treatment success was defined as reduction in number of FI episodes of ≥50% compared to baseline. Missing bowel diary data were considered treatment failures. The Wilcoxon signed rank test was used to compare changes observed at 12 months vs baseline. Results Mean age was 59.6 years old (SD 9.7). The mean duration of FI was 110 mo (range 8-712) months. Mean length of the implant procedure was 33.4 (SD 11.6) minutes. Mean EBL was 12.9 (SD 10.5) mL. Average follow-up was 24.9 months. At 12 months, 69.1% of patients met the criteria for treatment success, and 19% of subjects reported complete continence. FI episodes/wk decreased from a median of 9.0 (range 2-40) at baseline to 2.5 (range 0-40) ( P P P P P Conclusion The TOPAS system provides significant improvements in FI symptoms and quality of life with an acceptable AE profile and may therefore be a viable minimally invasive treatment option for FI in women.