Abstract

Due to the excessive contamination of heavy metals pollution, it is very urgent and necessary to develop a real-time detection method for the heavy metals in food. As a target sensing device, a paper-based microfluidic device (μPAD) has the advantages of simplicity, low-cost, and portability. In this study, a self-driven microfluidic paper-based chip was first developed for the simultaneous detection of four targets. The channels on the microfluidic chip were prepared by using wax printing and automatic screen printing on the filter paper, where liquid flowed by capillary force without pump assistance. Based on the specific binding ability of aptamers to heavy metals, a “turn-on” fluorescence aptasensor for the simultaneous detection of four heavy metal ions was developed on the proposed multi-channel device via smartphone imaging. The obtained fluorescence images were digitized into RGB color values by Image J software, and an M-mode was established to realize the quantitative detection of heavy metal ions. Under optimal conditions, the limits of detection for lead(II), mercury(II), cadmium(II), and arsenic(III) were 4.20 nM, 1.70 nM, 2.04 nM, and 1.65 nM, respectively. Furthermore, the aptasensor was successfully applied to the quantitative detection of four heavy metal ions in apple and lettuce samples with recovery rates of 84.0%–104.1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.