Abstract
In this study, we conducted a fully integrated point-of-care prothrombin time test on a microfluidic disk analyzer. The microfluidic functions integrated on the disk were capable of separating whole blood, decanting plasma, and mixing it with reagents in sequence under alternate spinning. The assay protocol was completed by alternate spinning without using microvalves or surface modification. Clinical sample tests on prothrombin time measurement were conducted by both the microfluidic disk analyzer and the reference instrument used in medical centers. The test results showed a good correlation and agreement between the two instruments.