Abstract

The design and first results from an electrically isolated and plasma-shielded Rogowski probe, used in the reconnection scaling experiment (RSX), are presented. The probe is designed to withstand extreme thermal shock, plasma corrosion, and be vacuum sanitary, which is accomplished with a machinable boron nitride shell. The novel miniature design, with an inner detecting area of 0.79 cm2, allows accurate position detection of plasma current channels with ≈2 cm radius and to measure local current density profiles. The temporal resolution (<1 μs) is sufficiently high to resolve the dynamic evolution of RSX plasma current channels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.