Abstract

Hydrogels that contain different amounts of amino groups were used to modify microcantilevers for the pH measurements. These microcantilevers deflected upon exposure to various pH solutions due to the swelling and shrinking of the hydrogels. The microcantilever deflection, as a function of pH, is nearly linear in a wide pH range. A significant 1000 nm/pH bending response was observed from a Gel‐4 coated microcantilever, which could be used for precise pH measurements. Such hydrogel coated microcantilevers could potentially be used to prepare microcantilever chemical and biological sensors when molecular recognition agents are immobilized into the polymer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.