Abstract

The results of Monte Carlo calculated dose distributions of proton treatment of ocular melanoma are presented. An efficient spot scanning method utilizing active energy modulation, which also minimizes the number of target spots was developed. We simulated various parameter values for the particle energy spread and the pencil beam diameter in order to determine values suitable for medical treatment. We found that a 2.5-mm-diameter proton beam with a 5% Gaussian energy spread was suitable for treatment of ocular melanoma while preserving vision for the typical case that we simulated. The energy spectra and the required proton current were also calculated and are reported. The results are intended to serve as a guideline for a new class of low-cost, compact accelerators.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.